Residual Solvent, USP 467 Class 2 MixA

| 1. Cyclohexane | 2. Methylcyclohexane | 3. trans-1,2-Dichloroethene + Tetrahydrofuran | 4. Methanol | 5. Methylene chloride |
| 6. cis-1,2-Dichloroethene | 7. Acetonitrile | 8. Toluene | 9. 1,4-Dioxane | 10. Ethylbenzene |
| 11. p-Xylene | 12. m-Xylene | 13. o-Xylene | 14. Chlorobenzene |
| Oven Temp. | 80℃ | Injection Time | 1 min | Vial Stirring | No | Needle Flush Time | 20 min |
|---|---|---|---|---|---|---|---|
| Thermostat Time | 60 min | Sample Line Temp. | 110℃ | Vial Pressure | 75 kPa | Transfer Line Temp. | 120℃ |
| Vial Pressurize Time | 1 min | Vial Volume | 20 mL | Loading Time | 0.5 min |
| Column | SH-PolarWax 30 m, 0.32 mm ID, 0.25 μm (P/N: 221-75972-30) | Makeup gas | 30 mL / min(He) |
|---|---|---|---|
| Column Temp. | 50℃(20 min)- 6℃ / min - 165℃(20 min) | Split ratio | 1:10 |
| Gas Velocity | 35 cm / sec(He) | H2 | 40 mL / min |
| FID Temp. | 250℃ | Air | 400 mL / min |