Residual Solvent

| Peak | tR (min) | Conc. (μg/mL) |
|---|---|---|
| 1. Acetonitrile | 2.056 | 410 |
| 2. Dichloromethane | 2.185 | 600 |
| 3. trans-1,2-Dichloroethene | 2.431 | 940 |
| 4. Hexane | 2.707 | 290 |
| 5. Nitromethane | 3.497 | 50 |
| 6. cis-1,2-Dichloroethene | 3.581 | 940 |
| 7. Tetrahydrofuran | 3.948 | 690 |
| 8. Chloroform | 4.125 | 60 |
| 9. 1,1,1-Trichloroethane | 4.329 | 1,500 |
| 10. Cyclohexane | 4.427 | 3,880 |
| Peak | tR (min) | Conc. (μg/mL) |
|---|---|---|
| 11. 2-Methoxyethanol | 5.079 | 50 |
| 12. Trichloroethene | 6.095 | 80 |
| 13. Methylcyclohexane | 6.373 | 1,180 |
| 14. 1,4-Dioxane | 6.554 | 380 |
| 15. 2-Ethoxyethanol | 6.982 | 160 |
| 16. Pyridine | 7.439 | 200 |
| 17. Toluene | 7.587 | 890 |
| 18. Ethylene glycol | 7.748 | 620 |
| 19. 2-Hexanone | 8.260 | 50 |
| 20. Dimethylformamide | 8.597 | 880 |
| Peak | tR (min) | Conc. (μg/mL) |
|---|---|---|
| 21. Chlorobenzene | 8.882 | 360 |
| 22. Ethylbenzene | 8.985 | 368 |
| 23. p-Xylene | 9.096 | 304 |
| 24. m-Xylene | 9.096 | 1,302 |
| 25. o-Xylene | 9.412 | 196 |
| 26. N,N-Dimethylacetamide | 9.654 | 1,090 |
| 27. 1-Methyl-2-pyrrolidone | 11.466 | 530 |
| 28. 1,2,3,4-Tetrahydronaphthalene | 12.116 | 100 |
| 29. Sulfolane | 13.266 | 160 |
| Column | SH-I-624Sil MS, 20m, 0.18mm ID, 1.00μm (P/N: 227-36075-01) | Flow Rate | 1.35 mL/min |
|---|---|---|---|
| Inj. Vol. | 0.5 μL split (split ratio 50:1) | Linear Velocity | 45 cm/sec |
| Liner | 5mm split | Dead Time | 0.74 min @ 40℃ |
| Inj. Temp. | 250℃ | Detector | FID @ 250℃ |
| Oven Temp. | 40℃ (hold 5 min) to 260℃ @ 50℃/min | Data Rate | 50 Hz |
| Carrier Gas | He, constant flow |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)