SHシリーズ
Inhalants

| Peaks | BAC Plus 1 tR (min) |
BAC Plus 2 tR (min) |
|---|---|---|
| 1. Methanol | 1.340 | 1.230 |
| 2. Ethanol | 1.694 | 1.470 |
| 3. Isopropanol | 2.072 | 1.660 |
| 4. Diethyl ether | 2.149 | 1.334 |
| 5. Acetone | 2.256 | 1.589 |
| 6. Methylene chloride | 2.405 | 1.703 |
| 7. MTBE | 3.185 | 1.660 |
| 8. Hexane | 3.683 | 1.660 |
| 9. Chloroform | 3.956 | 2.667 |
| 10. MEK | 4.210 | 2.543 |
| Peaks | BAC Plus 1 tR (min) |
BAC Plus 2 tR (min) |
|---|---|---|
| 11. Ethyl acetate | 4.524 | 2.414 |
| 12. Benzene | 5.390 | 2.883 |
| 13. Trichloroethylene | 6.306 | 3.522 |
| 14. MIBK | 8.101 | 5.775 |
| 15. Toluene | 8.162 | 5.440 |
| 16. Ethylbenzene | 10.286 | 7.681 |
| 17. m-Xylene | 10.487 | 7.879 |
| 18. p-Xylene | 10.487 | 7.879 |
| 19. o-Xylene | 10.990 | 8.467 |
| Columns | SH-BAC Plus 1 30m, 0.32mm ID, 1.80μm (P/N: 227-36260-01) and SH-BAC Plus 2 30m, 0.32mm ID, 0.60μm (P/N: 227-36263-01) using SH-I Guard Column (P/N: 227-36305-01) with universal “Y” Press-Tight connector |
Sample Equil. Time | 5 min |
|---|---|---|---|
| Diluent | Water | Vial Pressure | 30 psi |
| Conc. | 50 μg/mL each inhalant | Pressurize Time | 2 min |
| Injection | headspace-loop split (split ratio 50:1) | Loop Pressure | 20 psi |
| Liner | 1 mm straight inlet liner | Loop Fill Time | 1 min |
| Inj. Port Temp. | 220℃ | Oven Temp. | 40℃ (hold 4 min) to 120℃ at 10℃/min (hold 0 min) |
| Instrument | Tekmar HT3 | Carrier Gas | He, constant flow |
| Inj. Time | 3 min | Linear Velocity | 50 cm/sec |
| Transfer Line Temp. | 125℃ | Detector | FID @ 240℃ |
| Valve Oven Temp. | 125℃ | Make-up Gas Flow Rate | 30 mL/min |
| Sample Temp. | 70℃ | Make-up Gas Type | N2 |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)