SHシリーズ
Barbiturates《Acidic/Neutral Drugs》《Underivatized》

| 1. Barbital | 2. Aprobarbital | 3. Butalbital | 4. Amobarbital | 5. Talbutal | 6. Pentobarbital |
| 7. Methohexital | 8. Secobarbital | 9. Thiopental | 10. Hexobarbital | 11. Mephobarbital | 12. Phenobarbital |
| Column | SH-35, 30m, 0.32mm ID, 0.50μm (P/N: 227-36152-01) | Carrier Gas | He, constant pressure |
|---|---|---|---|
| Inj. Vol. | 1.0 μL split (split ratio 30:1) | Linear Velocity | 35 cm/sec @ 210℃ |
| Inj. Temp. | 300℃ | Detector | FID @ 300℃ |
| Oven Temp. | 210℃ (hold 2 min) to 300℃ @ 7℃/min (hold 2 min) | FID sensitivity | 5.12 x 10-10 AFS |
【出典】株式会社島津製作所SHシリーズアプリデータ集(株式会社島津ジーエルシー発行)