SHシリーズ
Propylene and C1-C5 Hydrocarbons

| 1. Methane | 2. Ethane | 3. Ethylene | 4. Propane | 5. Cyclopropane | 6. Propylene | 7. Isobutane |
| 8. n-Butane | 9. Propadiene | 10. Acetylene | 11. trans-2-Butene | 12. 1-Butene | 13. Isobutylene | 14. cis-2-Butene |
| 15. Isopentane | 16. n-Pentane | 17. 1,3-Butadiene | 18. Methylacetylene |
| Column | SH-Alumina BOND / Na2SO4, 50m, 0.53mm ID, 10μm (P/N:227-36301-01) | Carrier Gas | He, constant pressure (10 psi, 68.9 kPa) |
|---|---|---|---|
| Inj. Vol. | 10 μL split | Linear Velocity | 47 cm/sec @ 50℃ |
| Liner | 2 mm splitless | Detector | FID @ 200℃ |
| Inj. Temp. | 200℃ | Make-up Gas Type | N2 |
| Split Vent Flow Rate | 35 mL/min | Data Rate | 20 Hz |
| Oven Temp. | 50℃ to 150℃ @ 6℃/min |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)