SHシリーズ
Permanent Gases

Peaks
1. Ar (5.65 ppm) 2. CO (5.44 ppm) 3. O2 (5.54 ppm) 4. N2 (5.40 ppm) 5. CH4 (5.71 ppm)
| Instrument | GCMS-QP2010 Ultra | Control Mode | Pressure (100 kPa) |
|---|---|---|---|
| Column | SH-Msieve 5A, 30 m, 0.32 mm ID, 30 μm (P/N:227-36611-02) | Carrier Gas | Helium |
| Sample injection | Gas sampler (1 mL loop volume) (P/N: 223-57653-91) | Oven Temp | 35℃ (hold 2 min) to 150℃ @ 10℃/min (hold 5 min) |
| Inj. Mode | Split (split ratio: 50:1) | Detector | MS Interface Temp: 200℃ Ion Source Temp: 200℃ Measurement Mode: Scan (m/z 10 to100) Event Time: 0.5 sec Ionization Method: EI Emission Current:150 μA |
| Inj. Temp | 200℃ |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)