Fatty Acids《Free》

| 1. Acetic acid | 2. Propionic acid | 3. Isobutyric acid | 4. n-Butyric acid | 5. Isovaleric acid | 6. n-Valeric acid | 7. Isocaproic acid | 8. Caproic acid | 9. Heptanoic acid |
| 10. Caprylic acid | 11. Capric acid | 12. Lauric acid | 13. Myristic acid | 14. Palmitic acid | 15. Stearic acid | 16. Arachidic acid | 17. Behenic acid |
| Column | SH- PolarD, 30 m, 0.53 mm ID, 0.25 μm (P/N: 227-36278-01) | Carrier Gas | H2, constant flow |
|---|---|---|---|
| Conc. | 25 ng/μL | Linear Velocity | 80 cm/sec |
| Inj. Vol. | 0.3 μL Direct | Detector | FID @ 280℃ |
| Inj. Temp. | 280℃ | Flow rate | 10cc/min. |
| Oven Temp. | 100℃ (hold 2 min) to 280℃ @ 8℃/min (hold 10 min) | FID sensitivity | 128 x 10-11 AFS |