SHシリーズ
Fatty Acids《Free》

1. Acetic acid 2. Propionic acid 3. Isobutyric acid 4. n-Butyric acid 5. Isovaleric acid 6. n-Valeric acid 7. Isocaproic acid 8. Caproic acid 9. Heptanoic acid
| Column | SH-PolarD, 30m, 0.25mm ID, 0.25μm (P/N: 221-75981-30) | Carrier Gas | H2, constant pressure |
|---|---|---|---|
| Conc. | ~10-20 ng/μL | Linear Velocity | 40 cm/sec |
| Inj. Vol. | 1.0 μL split (split ratio 50:1) | Detector | FID @ 250℃ |
| Inj. Temp. | 250℃ | FID sensitivity | 2 x 10-11 AFS |
| Oven Temp. | 145℃ |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)