SHシリーズ
Flavor Volatiles

| 1. Hexane | 2. Acetaldehyde | 3. Acetone | 4. Methanol | 5. Ethyl acetate | 6. Isopropyl alcohol | 7. Ethanol | 8. Methylene chloride |
| 9. Ethyl propionate | 10. α-Pinene | 11. Ethylbutyrate | 12. n-Hexanal | 13. β-Pinene | 14. Sabinene | 15. Myrcene | 16. δ-Limonene |
| 17. 1,8-Cineole | 18. trans-2-Hexenal | 19. γ-Terpinene | 20. p-Cymene | 21. Terpinolene | 22. Octanal | 23. Nonanal | 24. cis-Limonene monoxide |
| 25. trans-Limonene | 26. Furfural | 27. Citronellal | 28. Decanal | 29. Linalool | 30. Octanol | 31. trans-α-Bergamotene | 32. β-Caryophyllene |
| 33. Terpinene-4-ol | 34. Neral | 35. α-Terpineol | 36. Neryl acetate | 37. Valencene | 38. Geranial | 39. Carvone | 40. Geranyl acetate |
| 41. d/l Carveol | 42. α-Ionone | 43. d/l Carveol | 44. β-Ionone |
| Column | SH-PolarWax, 60m, 0.53mm ID, 0.50μm (P/N: 227-36255-02) | Carrier Gas | He, constant linear velocity |
|---|---|---|---|
| Inj. Vol. | 0.8 μL split (split ratio 20:1) | Linear Velocity | 20 cm/sec @ 70℃ |
| Inj. Temp. | 220℃ | Detector | FID @ 260℃ |
| Oven Temp. | 70℃ (hold 15 min) to 190℃ @ 2℃/min (hold 5 min) | FID Sensitivity | 64 x 10-11 AFS |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)