SHシリーズ

| Peaks | tR (min) |
|---|---|
| 1. Ethanol | 4.18 |
| 2. Linalool* | 18.59 |
| 3. Terpineol | 21.21 |
| 4. Benzyl acetate | 21.78 |
| 5. Citronellol* | 22.29 |
| 6. Dipropylene glycol isomer 1 | 23.31 |
| 7. Geraniol* | 23.57 |
| Peaks | tR (min) |
|---|---|
| 8. Dipropylene glycol isomer 2 | 24.10 |
| 9. Dipropylene glycol isomer 3 | 24.20 |
| 10. Lilial* | 26.63 |
| 11. Isoeugenol* | 28.29 |
| 12. Piperonal | 29.20 |
| 13. Diethyl phthalate | 30.91 |
| 14. Benzyl salicylate* | 35.81 |
* Denotes fragrance allergen or suspected allergen
| Column | SH-PolarWax-MS, 30 m, 0.25 mm ID, 0.25 μm (P/N: 227-36322-01) |
Scan Program | Group | Start Time (min) |
Scan Range (amu) |
Scan Rate (scans/sec) |
|---|---|---|---|---|---|---|
| Conc. | Neat | 1 | 0 | 40-550 | 2 | |
| Inj. Vol. | 1 μL split (split ratio 200:1) | Transfer Line Temp. | 260℃ | |||
| Liner | Premium 3.5 mm ID single taper w/wool | Analyzer Type | Quadrupole | |||
| Inj. Temp. | 250℃ | Source Temp. | 250℃ | |||
| Oven Temp. | 35℃ (hold 5 min) to 250℃ at 7℃/min (hold 5 min) | Electron Energy | 70 eV | |||
| Carrier Gas | He, constant linear velocity | Solvent Delay Time | 0.5 min | |||
| Linear Velocity | 36 cm/sec | Ionization Mode | EI | |||
| Detector | MS | Instrument | Shimadzu 2010 GC & QP2010+ MS | |||
| Mode | Scan | |||||
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)