Volatile Organic Compounds

| 1. dichlorodifl uoromethane | 2. chloromethane | 3. vinyl chloride | 4. bromomethane | 5. chloroethane |
| 6. trichlorofl uoromethane | 7. 1,1-dichloroethene | 8. carbon disulfi de | 9. 1,1,2-trichloro-1,2,2-trifl uoroethane | 10. methylene chloride |
| 11. acetone | 12. trans-1,2-dichloroethene | 13. methyl acetate | 14. methyl tert-butyl ether | 15. 1,1-dichloroethane |
| 16. cis-1,2-dichoroethane | 17. cyclohexane | 18. bromochloromethane (IS) | 19. chloroform | 20. carbon tetrachloride |
| 21. 1,1,1-trichloroethane | 22. Propachlor | 23. benzene | 24. Cycloate | 25. 1,2-dichloroethane |
| 26. methylcyclohexane | 27. trichloroethene | 28. 1,4-difl uorobenzene (IS) | 29. 1,2-dichloropropane | 30. bromodichloromethane |
| 31. cis-1,3-dichloropropene | 32. toluene d8 (SS) | 33. toluene | 34. tetrachloroethane | 35. 4-methyl-2-pentanone |
| 36. trans-1,3-dichloropropane | 37. 1,1,2-trichloroethane | 38. dibromochloromethane | 39. 1,2-dibromoethane | 40. 2-hexanone |
| 41. chlorobenzene d5 (IS) | 42. chlorobenzene | 43. ethylbenzene | 44. m-xylene | 45. p-xylene |
| 46. o-xylene | 47. styrene | 48. bromoform | 49. isopropylbenzene | 50. 4-bromofl uorobenzene (SS) |
| 51. 1,1,2,2-tetrachloroethane | 52. 1,3-dichlorobenzene | 53. 1,4-dichlorobenzene | 54. 1,2-dichlorobenzene | 55. 1,2-dibromo-3-chloropropane |
| 56. 1,2,4-trichlorobenzene |
| Column | SH-VMS, 30 m, 0.25 mm ID, 1.40 μm (P/N: 227-36268-01) | Inj. | Split (split ratio: 35:1) |
|---|---|---|---|
| Purge and Trap | Trap:#10 (Tenax®/silica gel/carbon molecular sieve) | Inj. Temp | 200℃ |
| Sample Temp:ambient | Carrier Gas | He, linear velocity 34 cm/sec., 40℃, constant flow | |
| Purge:11 min @ 40 mL/min | Oven Temp | 40℃ (hold 4 min) to 90℃ @16 °C/min to 220℃ at 32℃/min (hold 5 min) | |
| Desorb preheat:185℃ | Detector | MS | |
| Desorb:0.5 min @ 190℃ | Transfer Line Temp: 150 °C | ||
| Desorb flow rate:35.0 mL/mi | Scan Range: 35-300 amu. | ||
| Bake:8 min @ 210℃ | Ionization: EI | ||
| Interface:split injector | |||
| Transfer Line Temp:150℃ |