SHシリーズ
PAHs

| Peaks | Quant Ion (m/z) | SIM Group # |
|---|---|---|
| 1. Naphthalene | 128 | 1 |
| 2. Biphenyl | 154 | 1 |
| 3. Acenaphthylene | 152 | 1 |
| 4. Acenaphthene | 153 | 1 |
| 5. Fluorene | 165 | 1 |
| 6. Dibenzothiophene | 184 | 2 |
| 7. Phenanthrene | 178 | 2 |
| 8. Anthracene | 178 | 2 |
| 9. 4H-Cyclopenta[def]phenanthrene | 190 | 2 |
| 10. Fluoranthene | 202 | 2 |
| 11. Pyrene | 202 | 2 |
| 12. Benzo[ghi]fluoranthene | 226 | 3 |
| 13. Benzo[c]phenanthrene | 228 | 3 |
| 14. Benz[a]anthracene | 228 | 3 |
| 15. Cyclopenta[cd]pyrene | 226 | 3 |
| 16. Triphenylene | 228 | 3 |
| 17. Chrysene | 228 | 3 |
| 18. Benzo[b]fluoranthene | 252 | 4 |
| Peaks | Quant Ion (m/z) | SIM Group # |
|---|---|---|
| 19. Benzo[k]fluoranthene | 252 | 4 |
| 20. Benzo[j]fluoranthene | 252 | 4 |
| 21. Benzo[a]fluoranthene | 252 | 4 |
| 22. Benzo[e]pyrene | 252 | 4 |
| 23. Benzo[a]pyrene | 252 | 4 |
| 24. Perylene | 252 | 4 |
| 25. Dibenz[a,j]anthracene | 278 | 5 |
| 26. Dibenz[a,c]anthracene | 278 | 5 |
| 27. Indeno[1,2,3-cd]pyrene | 276 | 5 |
| 28. Dibenz[a,h]anthracene | 278 | 5 |
| 29. Benzo[b]chrysene | 278 | 5 |
| 30. Picene | 278 | 5 |
| 31. Benzo[ghi]perylene | 276 | 5 |
| 32. Anthanthrene | 276 | 5 |
| 33. Dibenzo[b,k]fluoranthene | 302 | 6 |
| 34. Dibenzo[a,e]pyrene | 302 | 6 |
| 35. Coronene | 300 | 6 |
| 36. Dibenzo[a,h]pyrene | 302 | 6 |
| Column | SH-I-PAH, 40m, 0.18mm ID, 0.07μm (P/N:227-36073-01) | SIM Program | Group | Start Time (min) |
Ion(s) (m/z) | Dwell (ms) |
|---|---|---|---|---|---|---|
| Diluent | Toluene | 1 | 3.00 | 128, 152, 153, 154, 165 | 40 | |
| Conc. | 0.2 - 2 μg/mL (SRM 2260a PAH mix was diluted 5x in toluene) | 2 | 5.50 | 178, 184, 190, 202 | 50 | |
| Inj. Vol. | 0.5 μL pulsed splitless (hold 0.58 min) | 3 | 13.00 | 226, 228 | 100 | |
| Liner | Premium 2 mm single taper w/wool | 4 | 20.00 | 252 | 200 | |
| Inj. Temp. | 275℃ | 5 | 26.00 | 276, 278 | 100 | |
| Pulse Pressure | 80 psi (551.6kPa) | 6 | 29.00 | 300, 302 | 150 | |
| Pulse Time | 0.6 min | Transfer Line Temp. | 350℃ | |||
| Purge Flow | 40 mL/min | Analyzer Type | Quadrupole | |||
| Oven Temp. | 110℃ (hold 1 min) to 210℃ @ 37℃/min to 260℃ @ 3℃/min to 350℃ @ 11℃/min (hold 4.5 min) |
Source Temp. | 350℃ | |||
| Carrier Gas | He, constant flow | Quad Temp. | 200℃ | |||
| Flow Rate | 1.4 mL/min | Solvent Delay Time | 3.00 min | |||
| Detector | MS | Tune Type | PFTBA | |||
| Mode | SIM | Ionization Mode | EI | |||
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)