SHシリーズ
PAHs

| 1. Naphthalene | 2. 2-Methylnaphthalene | 3. 1-Methylnaphthalene | 4. Acenaphthylene | 5. Acenaphthene | 6. Fluorene |
| 7. Phenanthrene | 8. Anthracene | 9. Fluoranthene | 10. Pyrene | 11. Benz[a]anthracene | 12. Chrysene |
| 13. Triphenylene | 14. Benzo[b]fluoranthene | 15. Benzo[k]fluoranthene | 16. Benzo[j]fluoranthene | 17. Benzo[a]pyrene | 18. 3-Methylcholanthrene |
| 19. Dibenz[a,h]acridine | 20. Dibenz[a,j]acridine | 21. Indeno[1,2,3-cd]pyrene | 22. Dibenz[a,h]anthracene | 23. Benzo[ghi]perylene | 24. 7H-Dibenzo[c,g]carbazole |
| 25. Dibenzo[a,e]pyrene | 26. Dibenzo[a,i]pyrene | 27. Dibenzo[a,h]pyrene |
| Column | SH-I-17Sil MS, 30 m, 0.25 mm ID, 0.25 μm (P/N: 221-75916-30) | Oven Temp | 65℃ (hold 0.5 min) to 220℃ @ 15℃/min to 330℃ @ 4℃/min (hold 15 min) |
|---|---|---|---|
| Inj. Vol. | 0.5 μL splitless (hold 1.75 min) | Carrier Gas | He, constant flow rate 2.0 mL/min |
| Inj. Temp | 320℃ | Detector | FID, 320℃ |
| Purge Flow | 75 mL/min |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)