Organophosphorous Pesticides

| 1. Dichlorvos | 2. Hexamethylphosphoramide | 3. Mevinphos | 4. Trichlorfon | 5. TEPP | 6. Demeton-O | 7. Tributyl phosphate (SS) |
| 8. Thionazin | 9. Ethoprop | 10. Naled | 11. Sulfotepp | 12. Phorate | 13. Dicrotophos | 14. Monocrotophos |
| 15. Demeton-S | 16. Terbufos | 17. Dimethoate | 18. Diazinon | 19. Dioxathion | 20. Fonophos | 21. Disulfoton |
| 22. Phosphamidon isomer (breakdown product) |
23. Dichlorofenthion | 24. Phosphamidon | 25. Chlorpyrifos methyl | 26. Methyl parathion | 27. Ronnel | 28. Aspon |
| 29. Fenitrothion | 30. Malathion | 31. Chlorpyrifos | 32. Trichloronate | 33. Parathion-ethyl | 34. Fenthion | 35. Merphos |
| 36. Chlorfenvinphos | 37. Crotoxyphos | 38. Stirofos | 39. Prothiofos | 40. Merphos oxone (breakdown product) |
41. Ethion | 42. Fensulfothion |
| 43. Sulprofos | 44. Carbophenothion | 45. Famphur | 46. Triphenyl phosphate (SS) | 47. EPN | 48. Phosmet | 49. Leptophos |
| 50. Tri-o-cresyl phosphate | 51. Azinphos-methyl | 52. Azinphos-ethyl | 53. Coumaphos |
| Column | SH-OPP2, 30 m, 0.32 mm ID, 0.32 μm (P/N: 221-75887-30) | Carrier Gas | He |
|---|---|---|---|
| Inj. Vol. | 1 μL splitless (hold 1 min) | Dead Time | 1.03 min at 80℃ |
| Inj. Temp | 200℃ | Detector | FPD, 250℃ |
| Oven Temp | 80℃ (hold 0.5 min) to 280℃ @ 12℃/min (hold 10 min) | Notes | Constant pressure |