SHシリーズ
Sympathomimetic Amines《Basic Drugs》

| 1. Phenylethylamine | 2. Amphetamine | 3. Phentermine | 4. Methamphetamine | 5. Fenfluramine |
| 6. Mephentermine | 7. Cathinone | 8. Phenylpropanolamine | 9. Methcathinone | 10. Nicotine |
| 11. Ephedrine | 12. Pseudoephedrine | 13. Phenmetrazine | 14. Phendimetrazine | 15. Methylenedioxyamphetamine |
| 16. Diethylpropion | 17. Methylene dioxymethamphetamine | 18. Methylene dioxyethylamphetamine | 19. 4-Methyl-2,5-dimethoxyamphetamine | 20. Phenylephrine |
| 21. Caffeine | 22. Benzphetamine |
| Column | SH-200, 30m, 0.25mm ID, 0.50μm (P/N: 227-36181-01) | Carrier Gas | He |
|---|---|---|---|
| Inj. Vol. | 1 μL split | Detector | MS |
| Inj. Temp | 250℃ | Mode | Scan |
| Split Vent Flow Rate | 45 mL/min | Ionization Mod | EI |
| Oven Temp. | 100℃ to 310℃ @ 10℃/min | Scan Range | 40-600 amu |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)