SHシリーズ
Hexamethylenediamine

| 1. cyclohexane | 2. hexamethyleneimine | 3. 1,4-diaminobutane | 4. pentamethylenediamine | 5. 1,2-diaminocyclohexane |
| 6. 1,5-diamino-2-methylpentane | 7. aminomethylcyclopentylamine | 8. hexamethylenediamine | 9. 6-aminocapronitrile | 10. n-valeramide |
| 11. adiponitrile | 12. bis-hexamethylenetriamine |
| Column | SH-PolarX, 30 m, 0.53 mm ID, 0.5 μm (P/N:227-36299-01) | Carrier gas | hydrogen |
|---|---|---|---|
| Oven temp. | 95 °C (hold 6 min.) to 235 °C @ 7 °C/min. (hold 2 min.) | Linear velocity | 40 cm/sec. (flow rate:5 cc/min.) |
| Inj./det. temp. | 255 °C | FID sensitivity | 64 x 10-11 AFS |
【出典】株式会社島津製作所SHシリーズアプリデータ集 (株式会社島津ジーエルシー発行)